Typical Installation
Replace existing station with Crest replacement, referencing the wiring table below.
When installing ensure the o-rings and gasket are correctly used to ensure a waterproof seal.
Always test all nurse call stations for correct operation (including cord out if applicable) before placing into service.
P/N 9A3006C
P/N 9A3006CC
Shower Pullcord Station,
Crest Replacement for Dukane
Technical Specifications
Electrical Rating: 24 VDC @ 20 mA
Mounting: Screw to standard single gang
wall box
Terminals: Wire Leads
Plate Size: 4.5"H x 2.75"W (1-gang)
Crest Technical Support: 1-800-328-8908
Specications subject to change without notice.
Visit CrestHealthcare.com for additional technical documents.
© Crest Healthcare 2019 B-39169 27 August 2019 / Rev. 03
qual ity re p lac eme nt
9A3006C
9A3006CC
Function Dukane Crest Pin #
EXTSENS White White 1
SUPERVISION Black Black 2
2CLASENS Green Green 3
VAN Brown Brown 4
CLWINHIBIT Blue / White Blue / White 5
FLASH Yellow Yellow 6
CLA Violet Violet 7
GROUND Black Black 8
+VL Red Red 9
EXCLR Blue Blue 10